RESOURCES >
REF ELECTRODES > LUGGIN PROBE
A
while ago I
got a question about constructing a Luggin capillary (Luggin probe or Luggin-Haber capillary) for a reference electrode. My answer formed the
basis of the notes below.

A Luggin capillary is a way of effectively bringing the potential sensing tip of the
reference electrode close to the surface of the working electrode, thereby
reducing the uncompensated iR drop. The Luggin probe may be open at the bottom
end and filled with the test electrolyte. It is often closed at the end with a
Vycor® frit and filled with a salt solution to form a salt-bridge. A Luggin
is fairly well described on the
Gamry website.
Since the goal of using a Luggin is to bring a narrow outside
diameter capillary close to your working electrode, most manufacturers will
design a Luggin to be compatible with the geometry of their cells. Your best
route is either to order one from the manufacturer of your cells, or to make one
yourself.
The design criteria are:
The outside diameter of the tip should be small. ASTM
G5 recommends that a Luggin tip be no closer than two tip diameters or so from the
surface of the working electrode (Ref 2). Otherwise the tip will shield part of the
working electrode and disturb the current flow path. So, the smaller the tip,
the closer you can get to the electrode surface. This helps to minimize the
uncompensated resistance of the cell setup.
Unfortunately, if the outside diameter is small,
the inside diameter is smaller! This leads to a high resistance in the Luggin/salt
bridge. See the next section.
Keep the
length of the narrow capillary short. Otherwise the
resistance of the Luggin can be high, even if it is filled with a very
conducting solution! At one time I measured the resistance of a Luggin filled
with 0.1M salt and it was about 100 Kohm. High resistances in the reference
electrode path can cause problems by causing a potentiostat to
oscillate. Many
commercial potentiostats have problems with reference electrode impedances above
about 10 Kohm.
The problem of high resistance salt bridges or
Luggin probes can be overcome in several ways. One (Ref 3) is
to simply short out the high resistance with a thin piece of platinum wire
placed in the narrow section of the capillary. No external connection to the
platinum is required. Another (Ref 4,
Ref 5) is to place a piece of platinum close to the tip of
the Luggin. The platinum is coupled to the reference electrode input of the potentiostat using a 0.1-0.01 µF
capacitor. This has the extra advantage of couteracting the effects of a high
resistance reference electrode as well as the effects of a high resistance
capillary.
|